Guided workflow

Add a Scale Bar to a Microscopy Image

Calibrate your microscopy image, add a readable physical scale bar, and export a research figure using NotyScale. Follow the steps below with an SEM, TEM or optical microscopy image and a verified scale reference.

  • NotyScale
  • 6 steps

Method NotyScale

Start with Microscopy image + verified scale reference

  1. Open Your Image
  2. Verify Calibration
  3. Prepare the Image
  4. Add Scale Bar
  5. Adjust Appearance
  6. Export and Verify

Result Exported image with an accurate, readable scale bar

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

A verified scale reference is required to create an accurate physical scale bar. Pixel dimensions or a magnification label alone cannot establish a reliable real-world distance.

  • Use the original microscopy image, or a copy whose resizing and resampling history is known.
  • Have an existing bar with a known length, an image-specific calibration, or another verified reference distance.
  • Confirm the physical unit, such as nm or ?m. Keep the original image and its calibration record.

Your workflow

Step-by-step workflow

  1. Open Your Image

    Open NotyScale in Workflow mode. Use the Image Library or the upload action to load your microscopy image, then confirm that the intended image is active. Keep an unmodified copy of the source image.

    The application groups the task into Import, Calibrate, and Scale & Export. This page breaks those stages into six practical checks.

  2. Verify Calibration

    Confirm that the attached calibration belongs to this exact image and acquisition scale. If no verified calibration is attached, open the Calibrate stage and use the manual two-point calibration control.

    1. Mark the two endpoints of a trustworthy reference distance.
    2. Continue to enter the corresponding real-world length and its unit. For example, enter 500 nm only if the reference actually represents 500 nm.
    3. Save the calibration and check that physical units are available before continuing.
    NotyScale workflow showing an attached calibration and Image Preparation controls beside a microscopy image
    Check the attached image calibration before adding a new physical scale bar.
  3. Prepare the Image

    If a crop is needed, open Image Preparation before adding the final bar. Calibrate first when the original bar is your only reference, then prepare the region of interest without hiding scientifically relevant features.

    Cropping without resampling preserves the physical distance per pixel. Resizing or resampling requires a calibration check before you continue.

  4. Add Scale Bar

    In Scale & Export, select Add Scale Bar. If your version offers an Auto Create size choice, Publication provides a heavier starting style for a small figure panel.

    Set Length to a useful physical distance in the verified unit. NotyScale calculates the pixel length from the calibration. Choose a value appropriate to the field of view, rather than selecting a physical length only for its appearance.

  5. Adjust Appearance

    Use the Layout and Label controls to refine the bar. In versions that show Length above the style tabs, set the physical length there, then use Layout for placement. Preview the result at approximately its final manuscript or presentation size.

    • Place the bar away from structures needed to interpret the image.
    • Choose sufficient thickness, readable label size and an unambiguous unit.
    • Adjust color or background contrast so the bar and text remain clear.
    NotyScale Layout panel showing scale-bar length, position and thickness controls
    Refine the physical length and position, then check label readability at the final figure size. Control placement can vary by application version.
  6. Export and Verify

    Select Export Image and choose the available format and resolution. Free PNG and JPEG exports are available; sign in to export or save your result.

    Open the saved file and complete the quality check below. Keep the original image and calibration alongside the exported figure. Compare the example below to see how the annotation changes the presentation; verify the calibration of your own image independently.

    SEM field before adding a scale bar in NotyScale
    BeforeAfter
    The same SEM field before and after adding a scale bar in NotyScale. Each source image needs its own verified calibration.

If your image already has a scale bar

Keep an existing bar if it is accurate, readable and appropriately positioned. If you replace it, verify its reference before cropping it out and avoid leaving conflicting annotations. Prefer an original, unannotated image when available; preserve the underlying scientific evidence.

Common problems and what to check

  • Scale-bar creation is unavailable: check that the correct image is active and that a valid calibration is attached.
  • The length looks incorrect: check the known reference, its unit, and any resizing or resampling after calibration.
  • The bar is hard to read: adjust thickness, label size, position or contrast, then review at the final figure size.
  • No scale information is available: check the acquisition files, instrument metadata or laboratory records. Contact the imaging facility before assigning a physical scale.

Final quality check

  • Accuracy: the physical length and unit agree with a valid calibration for this exact image.
  • Readability: the bar and numeric label remain clear at the final display size.
  • Integrity: the aspect ratio is preserved and important specimen features remain visible.
  • Reproducibility: retain the source image, calibration record and exported file together.

Start Scale Bar Workflow

Prepare your input and follow the guided steps in NotyScale.