Documentation
Microscopy Scale Bar Pre-Publication Checklist
Check calibration evidence, scale bar units, final figure readability and export integrity before submitting a microscopy figure.
- Applies to NotyScale 2.1.0.5
- Updated
When to use this checklist
Use this checklist after creating your scale bar and assembling the intended figure. It is a final verification record: the settings reference explains the controls, while this page asks what evidence confirms the finished figure is correct. Retain the source image, calibration reference and exported result so a failed check can be traced and repaired.
1. Verify the source and calibration evidence
Pass when the file being submitted can be traced to the original acquisition and the calibration describes the pixel scale of the image used for the bar. Record the reference distance, unit and pixel endpoints or the trusted pixel-size metadata. Similar dimensions, appearance or magnification labels are insufficient grounds for copying a calibration.
Example: a verified 20 micrometer reference spanning 400 original pixels gives 0.05 micrometers per pixel. A 10 micrometer bar should span 200 of those pixels. If the working image was resampled, repeat this check with the actual transformation; do not silently reuse the original pixel ratio.
2. Confirm the displayed distance and unit
Pass when the numeric label, unit and rendered bar agree with the verified calibration. Inspect the active image, not a previous image in the session. Check for nm versus micrometers versus millimeters and for px labels accidentally presented as physical measurements. If the figure contains several panels, record this check for each independently calibrated image.
3. Review image processing and crop boundaries
Pass when the figure preserves the scientific information needed for interpretation, any relevant processing is recorded and a bar or label has not been clipped. Keep the unmodified source, including any original microscope footer. A crop of unchanged pixels and a resampling operation have different effects on pixel scale. Resolve uncertainty about which operation occurred before approving the bar.
4. Inspect the final figure at submission size
Pass when the value and unit remain legible in the actual manuscript or poster layout. Inspect the line, end marks or ruler ticks at that size, on both light and dark areas as applicable. Confirm that the overlay covers no feature used in the analysis. If a label is hidden, verify that its physical meaning is stated unambiguously elsewhere in the figure or caption.
5. Check the file delivered to the publisher
Pass when the exported file opens in the receiving application, matches the intended pixel dimensions and print requirements and contains the expected bar. Recheck after any compression, conversion or figure assembly. A change to DPI alone does not add source detail. If later editing stretches the image independently of the bar, return to a proportional version and verify the result again.
Record the sign-off
- Record the source filename, calibration reference and unit for each panel.
- Record crop or resampling operations and the final output dimensions and DPI where relevant.
- Open the delivered file and record whether scale, legibility, margins and image integrity pass.
- Resolve every failed check before submission; retain the checked export with its source and calibration.
Continue your work
- Documentation NotyScale Right Panel: Complete Settings Reference Understand every section of the NotyScale right panel: calibration, length, presets, Style, Text, Color, Layout and export controls.
- Documentation NotyScale Documentation: Getting Started Learn the NotyScale workspace, image import options, guided workflow and session saving.
- Documentation NotyScale Calibration and Pixel Size Select, verify and change an image calibration before creating a physical scale bar.
- Documentation Prepare and Crop Images in NotyScale Use Image Preparation to crop or clean an image and review the result before adding a scale bar.
- Guide How to Add a Scale Bar to an SEM Image Online Without ImageJ Use the guided NotyScale Workflow to load a SEM image, manually calibrate it, optionally prepare the image, add a publication scale bar, set its length, and open export.