Guided workflow
Review Combined TGA and DSC Events
Review simultaneous or aligned TGA–DSC signals together so thermal events can be compared with mass changes over the same temperature region. The workflow helps establish temporal or temperature coincidence while keeping chemical interpretation separate from the measured evidence.
- NotyPlot
- TGA DSC
- 6 steps
Start with Combined TGA-DSC data with identifiable temperature, mass and heat-flow signals.
- Import both signals
- Verify channel alignment
- Analyze thermal events
- Compare event ranges
- Review the interpretation
- Report aligned evidence
Result A joint review of thermal events and mass-loss behavior.
Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.
Before you start
- Combined or alignable TGA and DSC data with a common temperature or time basis.
- Correct mass and heat-flow channel mapping and units.
- Known heat-flow sign convention, atmosphere and heating program.
Your workflow
Step-by-step workflow
Import both signals
Import the combined data and map the temperature, mass and heat-flow channels correctly.
Verify channel alignment
Confirm that the TGA and DSC channels share the intended temperature/time alignment.
Analyze thermal events
Run thermal analysis to review TGA mass-loss steps and DSC thermal events.
Compare event ranges
Compare event ranges, DTG peaks, DSC onset/peak/endset positions and residual mass behavior.
Review the interpretation
Separate observations of coincidence from proposed chemical mechanisms.
Report aligned evidence
Report the aligned evidence with the heating program, atmosphere and signal conventions.
Final quality check
- TGA and DSC channels are actually aligned before event temperatures are compared.
- Mass-loss steps and DSC events are reported as measured relationships rather than automatic causal assignments.
- Heat-flow direction is interpreted correctly.
- Overlapping events and shoulders are acknowledged.
- Use the same temperature calibration or alignment basis for both channels.
- Report the simultaneous-analysis test conditions.
- Support reaction assignments with chemistry, evolved-gas analysis or other evidence when necessary.
Scientific notes
- Coincident thermal and mass changes do not by themselves identify the reaction mechanism.
- Different channel response times or misalignment can make events appear shifted.
- Complex overlapping processes may not separate cleanly into one DSC event and one TGA step.
Common questions
What can combined TGA–DSC show?
It can show whether heat-flow events and mass changes occur in the same temperature region, which helps narrow interpretation.
Does an endotherm with mass loss prove evaporation?
No. It is consistent with several possibilities. Use sample chemistry and other evidence before assigning a mechanism.
Why do TGA and DSC event temperatures differ slightly?
Real kinetic differences, signal response and channel alignment can all contribute. Verify the shared temperature basis before interpreting the offset.
Related resources
- Workflow Analyze TGA Mass Loss and DTG Analyze thermogravimetric mass-loss steps together with the derivative thermogravimetry (DTG) curve. NotyPlot uses the derivative to help resolve loss events and reports step ranges, DTG peak temperatures and residual mass for review.
- Workflow Automatic XRD Peak Analysis in NotyPlot Run a guided automatic XRD peak-analysis pass, then review the detected and fitted results rather than treating automation as a final interpretation. The workflow is designed to shorten setup while keeping peak selection, fitting and scientific checks visible.
- Workflow Calculate Crystallite Size and Microstrain with Williamson-Hall Separate size-related and strain-related contributions to XRD peak broadening with a Williamson–Hall analysis built from several reviewed fitted peaks. The workflow emphasizes peak eligibility and the transformed linear fit because a numerical intercept and slope are only useful when the selected peaks support the model.
- Workflow Calculate Crystallite Size from XRD Using the Scherrer Equation Estimate coherent crystallite size from suitable fitted XRD peak widths using the Scherrer equation. The workflow keeps wavelength, shape factor, fitted FWHM and instrumental broadening visible because each directly affects the reported size.
Start this workflow
Prepare your input and follow the guided steps in NotyPlot.