Calculate Crystallite Size from XRD Using the Scherrer Equation

Estimate coherent crystallite size from suitable fitted XRD peak widths using the Scherrer equation. The workflow keeps wavelength, shape factor, fitted FWHM and instrumental broadening visible because each directly affects the reported size.

  • NotyPlot
  • XRD
  • 6 steps

Method NotyPlot

Start with XRD peaks with known wavelength and appropriate peak-width information.

  1. Verify the pattern and wavelength
  2. Detect and fit suitable peaks
  3. Review the fitted widths
  4. Set the broadening model
  5. Calculate crystallite size
  6. Report the assumptions

Result Scherrer crystallite-size estimates for reviewed peaks.

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

  • One or more suitable fitted XRD peaks with reliable FWHM.
  • The X-ray wavelength used for the scan.
  • A chosen Scherrer shape factor K and, when available, instrumental broadening information.

Your workflow

Step-by-step workflow

  1. Verify the pattern and wavelength

    Import the XRD pattern and verify wavelength, angle units and baseline state.

  2. Detect and fit suitable peaks

    Detect and fit suitable isolated or defensibly resolved peaks.

  3. Review the fitted widths

    Select the peaks to use and review the fitted FWHM values.

  4. Set the broadening model

    Set the Scherrer shape factor and choose the appropriate instrumental-broadening correction model or profile inputs.

  5. Calculate crystallite size

    Run the crystallite-size calculation and compare estimates across the selected peaks.

  6. Report the assumptions

    Report the reviewed estimate together with wavelength, K, peak selection and broadening assumptions.

Final quality check

  • Only reliable fitted widths are used.
  • Instrumental broadening is handled consistently with the available instrument information.
  • Peak-specific estimates do not vary wildly without investigation.
  • The result is described as crystallite or coherent-domain size, not automatically as particle size.
  • Record K, wavelength, fitted peak model and broadening correction.
  • Use consistent settings when comparing samples.
  • Investigate peaks affected by overlap, anisotropy or strong asymmetry before including them.

Scientific notes

  • Scherrer size is a model-based estimate of coherent diffracting domains and may differ substantially from microscopy particle size.
  • Peak broadening can contain strain, defects and instrumental contributions in addition to finite crystallite size.
  • Results become unstable when corrected broadening is very small or the fitted width is unreliable.

Common questions

Is Scherrer crystallite size the same as particle size?

No. Scherrer estimates coherent diffracting-domain size from peak broadening; a particle can contain one or more crystallites.

Do I need instrumental broadening correction?

If you want a physically meaningful size estimate and instrument broadening is not negligible, it should be considered using appropriate instrument information.

Can I use one XRD peak?

A single suitable peak can provide an estimate, but comparing several reliable peaks helps reveal anisotropy, overlap or fitting problems.