Guided workflow

Calculate Crystallite Size and Microstrain with Williamson-Hall

Separate size-related and strain-related contributions to XRD peak broadening with a Williamson–Hall analysis built from several reviewed fitted peaks. The workflow emphasizes peak eligibility and the transformed linear fit because a numerical intercept and slope are only useful when the selected peaks support the model.

  • NotyPlot
  • XRD
  • 6 steps

Method NotyPlot

Start with Several suitable XRD peaks with known wavelength and corrected broadening.

  1. Verify wavelength and broadening
  2. Detect and fit suitable peaks
  3. Select reliable fitted widths
  4. Inspect the transformed fit
  5. Review scatter and warnings
  6. Report size and microstrain

Result A Williamson-Hall fit with size and strain estimates.

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

  • Several fitted XRD peaks with reliable corrected widths.
  • Known X-ray wavelength and appropriate instrumental-broadening information.
  • Peak selection covering a useful angular range for the Williamson–Hall fit.

Your workflow

Step-by-step workflow

  1. Verify wavelength and broadening

    Import the XRD pattern and confirm wavelength and instrument-broadening settings.

  2. Detect and fit suitable peaks

    Detect and fit a set of suitable peaks using a consistent fitting approach.

  3. Select reliable fitted widths

    Select only peaks with defensible fitted widths for Williamson–Hall analysis.

  4. Inspect the transformed fit

    Run the transformation and inspect the included points and fitted line in the Williamson–Hall view.

  5. Review scatter and warnings

    Review scatter, slope, intercept and warnings; remove a peak only for a defensible fit or data-quality reason.

  6. Report size and microstrain

    Report size and microstrain estimates together with the model, selected peaks and fit limitations.

Final quality check

  • Enough suitable peaks are included to support a linear trend.
  • No single questionable peak dominates the fitted slope or intercept.
  • Instrument-corrected widths remain physically meaningful.
  • Scatter and fit quality are acknowledged rather than hidden by the final values.
  • Keep a record of included and excluded peaks.
  • Use the same width definition and correction settings across all peaks.
  • Describe the specific Williamson–Hall model assumption when interpreting size and strain.

Scientific notes

  • Williamson–Hall assumes a simplified relationship between size and strain broadening; real materials may show anisotropy or nonuniform strain that violates it.
  • A small number of peaks or strong scatter can make the fit unstable.
  • Instrumental broadening and peak overlap can bias both the size and microstrain estimates.

Common questions

How many peaks do I need for Williamson–Hall analysis?

Use several reliable fitted peaks across a useful angular range. More points do not help if their FWHM values are poor or inconsistent.

Why is the Williamson–Hall plot scattered?

Possible causes include peak overlap, anisotropic broadening, poor fits, instrument correction problems or a material that does not follow the simplified model.

Is Williamson–Hall better than Scherrer?

They answer related but different questions. Scherrer gives a size estimate from peak width, while Williamson–Hall uses several peaks to model size and strain contributions under additional assumptions.

Start this workflow

Prepare your input and follow the guided steps in NotyPlot.