Guided workflow

Analyze TGA Mass Loss and DTG

Analyze thermogravimetric mass-loss steps together with the derivative thermogravimetry (DTG) curve. NotyPlot uses the derivative to help resolve loss events and reports step ranges, DTG peak temperatures and residual mass for review.

  • NotyPlot
  • TGA
  • 6 steps

Method NotyPlot

Start with TGA data with temperature and mass or mass-percentage values.

  1. Import the TGA data
  2. Verify mass normalization
  3. Calculate DTG and mass-loss steps
  4. Review step ranges and peaks
  5. Inspect the residual mass
  6. Report the test conditions

Result Reviewed mass-loss steps and a DTG curve.

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

  • TGA data with temperature and mass or mass-percentage values.
  • Correct mass normalization and temperature direction.
  • Relevant atmosphere, heating rate and sample conditions for interpretation.

Your workflow

Step-by-step workflow

  1. Import the TGA data

    Import the TGA dataset and verify temperature and mass columns and units.

  2. Verify mass normalization

    Inspect the starting mass or normalization and confirm that the mass curve is physically plausible.

  3. Calculate DTG and mass-loss steps

    Run thermal analysis to calculate the DTG curve and detect mass-loss steps.

  4. Review step ranges and peaks

    Review each step range, mass-loss percentage and DTG peak temperature against the original mass curve.

  5. Inspect the residual mass

    Inspect the residual mass and check whether small or overlapping events were missed or merged.

  6. Report the test conditions

    Report reviewed mass-loss intervals and residual mass with the test conditions and any smoothing used for derivative review.

Final quality check

  • DTG features coincide with real slope changes in the mass curve.
  • Mass-loss percentages are based on the intended starting mass or normalization.
  • Residual mass is plausible for the sample and test endpoint.
  • Minor events near the noise floor are not over-interpreted.
  • State atmosphere and heating rate when comparing or interpreting decomposition temperatures.
  • Keep the original mass curve visible alongside DTG.
  • Do not assign chemical reactions solely from mass-loss temperature without sample context or complementary evidence.

Scientific notes

  • Differentiation amplifies noise, so DTG shape depends on signal quality and smoothing.
  • Overlapping reactions may appear as shoulders or merged derivative peaks.
  • The automatic threshold may not capture very small mass-loss events.

Common questions

What is DTG used for?

DTG shows the rate of mass change and can make overlapping or subtle TGA steps easier to locate than the mass curve alone.

Why is a small TGA step missing?

Very small events may fall below the automatic detection threshold or be obscured by noise and overlap. Review the original curve and DTG together.

Does a DTG peak identify a decomposition reaction?

No. It locates a rate maximum; chemical assignment requires sample context and often complementary analysis.