Guided workflow

Review DSC Thermal Transitions

Detect and review DSC thermal events using the measured heat-flow curve, including peak position and extrapolated onset/endset information where the event can be resolved. The workflow keeps heat-flow direction and baseline context visible because those choices affect interpretation.

  • NotyPlot
  • DSC
  • 6 steps

Method NotyPlot

Start with DSC data with temperature and heat-flow information.

  1. Import the DSC data
  2. Confirm heat-flow direction
  3. Analyze thermal events
  4. Review event temperatures
  5. Check missed events
  6. Report reviewed transitions

Result Reviewed thermal transition events and a plotted DSC curve.

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

  • DSC data with temperature and heat-flow values.
  • Knowledge of the instrument heat-flow sign convention and heating/cooling direction.
  • Relevant test conditions such as heating rate, atmosphere and sample history.

Your workflow

Step-by-step workflow

  1. Import the DSC data

    Import the DSC dataset and verify temperature and heat-flow columns and units.

  2. Confirm heat-flow direction

    Confirm whether upward or downward heat-flow excursions represent endothermic or exothermic behavior for the source data.

  3. Analyze thermal events

    Inspect the baseline and run thermal-event analysis.

  4. Review event temperatures

    Review detected events, including onset, peak, endset and integrated area where available.

  5. Check missed events

    Add or correct an event manually only when the measured curve supports it and automatic detection missed the region.

  6. Report reviewed transitions

    Report reviewed event temperatures and areas together with the measurement conditions and sign convention.

Final quality check

  • Detected events correspond to visible curve features.
  • Onset and endset positions fall on plausible event flanks.
  • Integrated area is interpreted in the context of the Y-axis units rather than automatically called enthalpy in absolute units.
  • Heating/cooling direction and heat-flow convention are clear.
  • Report heating rate, atmosphere and sample mass/normalization when relevant.
  • Do not assign glass transition, melting, crystallization or reaction identity from curve shape alone without supporting context.
  • Use comparable test conditions when comparing samples.

Scientific notes

  • Overlapping transitions may not be separable by automatic event detection.
  • Baseline choice strongly influences onset and integrated area.
  • Thermal-event direction depends on instrument convention and may need interpretation before labeling endothermic or exothermic behavior.

Common questions

What does DSC onset mean here?

The workflow uses an extrapolated-onset style construction where possible, based on the event flank and local baseline rather than the first visible deviation.

Is the peak area automatically an enthalpy in J/g?

Only if the input signal and normalization support that unit. Treat the integrated area according to the actual Y-axis units and test metadata.

Can NotyPlot identify what caused a DSC peak?

It can detect and measure events, but assigning a physical or chemical mechanism requires sample knowledge and often complementary evidence.

Start this workflow

Prepare your input and follow the guided steps in NotyPlot.