Guided workflow

Find Raman Peaks

Detect Raman spectral peaks and review them against the measured curve before using positions for material comparison or interpretation. The workflow emphasizes baseline and fluorescence because they can dominate weak Raman features.

  • NotyPlot
  • RAMAN
  • 6 steps

Method NotyPlot

Start with A Raman intensity spectrum with shift units.

  1. Import the Raman spectrum
  2. Inspect background and noise
  3. Set detection sensitivity
  4. Detect and inspect peaks
  5. Review shoulders and overlap
  6. Export reviewed peaks

Result Reviewed Raman peak features.

Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.

Before you start

  • A Raman spectrum with a known Raman-shift axis and intensity signal.
  • Sufficient spectral resolution and signal-to-noise for the features of interest.
  • Awareness of strong fluorescence background or cosmic-ray artifacts when present.

Your workflow

Step-by-step workflow

  1. Import the Raman spectrum

    Import the Raman spectrum and verify the shift and intensity columns.

  2. Inspect background and noise

    Inspect fluorescence background, spikes and noise before running peak detection.

  3. Set detection sensitivity

    Choose a suitable sensitivity and apply baseline processing only when justified.

  4. Detect and inspect peaks

    Detect peaks and compare each important candidate with the original spectrum.

  5. Review shoulders and overlap

    Review closely spaced or shoulder features before treating them as separate peaks.

  6. Export reviewed peaks

    Keep the reviewed peak list and annotate or export only the features relevant to the analysis.

Final quality check

  • Detected peaks correspond to real spectral features rather than spikes or background variation.
  • Peak positions use the correct shift unit.
  • Close features are not over-resolved by sensitivity alone.
  • Baseline treatment does not remove broad Raman bands.
  • Keep processing settings with reported peak positions.
  • Use reference spectra or literature for identification rather than relying on position alone.
  • Compare samples using consistent processing when possible.

Scientific notes

  • Fluorescence can obscure weak Raman bands and distort peak detection.
  • Cosmic rays or isolated spikes can look like narrow peaks.
  • Peak position alone is not a unique material identification.

Common questions

Why does Raman peak detection find spikes?

Very narrow artifacts can resemble peaks. Inspect the raw spectrum and remove only known artifacts using a documented rule.

Can I remove fluorescence background?

Baseline processing can help, but verify that broad real bands are preserved and report the processing used.

Does a matching Raman shift prove a material identity?

No. Use the full pattern, relative features, sample context and reference data.

Start this workflow

Prepare your input and follow the guided steps in NotyPlot.