Guided workflow
Review FTIR Functional Groups
Review possible FTIR functional groups using detected band evidence, not single-peak lookup alone. NotyPlot links suggestions to measured spectral features so you can consider supporting bands, overlaps and sample chemistry before retaining an interpretation.
- NotyPlot
- FTIR
- 6 steps
Start with An FTIR spectrum and known sample context.
- Prepare a reviewed band list
- Inspect supporting bands
- Check related features
- Consider the sample chemistry
- Review ambiguous assignments
- Export supported annotations
Result Evidence-linked functional-group suggestions for review.
Figure 1. The workflow at a glance: what you start with, the steps, and what you leave with.
Before you start
- An FTIR spectrum with correctly interpreted signal orientation and wavenumber axis.
- Reviewed detected bands.
- Basic sample context such as composition, treatment, expected chemistry or known contaminants.
Your workflow
Step-by-step workflow
Prepare a reviewed band list
Import and analyze the FTIR spectrum to create a reviewed band list.
Inspect supporting bands
Open the functional-group evidence and inspect which measured bands support each suggestion.
Check related features
Check whether related expected bands are present and whether important contradictory features exist.
Consider the sample chemistry
Consider overlap, mixtures, baseline artifacts and sample chemistry before accepting a suggestion.
Review ambiguous assignments
Retain supported interpretations and mark ambiguous cases rather than converting every suggestion into a definitive assignment.
Export supported annotations
Export or annotate the spectrum with only the assignments needed for the scientific discussion.
Final quality check
- Assignments are supported by more than an isolated approximate band match when possible.
- Suggested groups are chemically plausible for the sample.
- Overlapping ranges and alternative assignments are acknowledged.
- The final labels remain traceable to measured band evidence.
- Use the interpretation as supporting evidence alongside complementary characterization when needed.
- Report important processing choices and spectrum type.
- Avoid naming a specific compound from generic functional-group ranges alone.
Scientific notes
- Functional-group ranges overlap and are not unique fingerprints by themselves.
- Mixtures, hydrogen bonding and local chemical environment can shift or broaden bands.
- Knowledge-base suggestions support review; they do not replace reference spectra or expert chemical interpretation.
Common questions
Can FTIR identify a compound from one band?
Usually not. Single bands are often non-unique; use multiple supporting features, sample context and reference information.
Why do several functional groups match the same band?
Vibrational ranges overlap. The useful question is which assignment is supported by the complete spectrum and sample chemistry.
Are NotyPlot functional-group suggestions final identifications?
No. They are evidence-linked suggestions for review.
Related resources
- Workflow Analyze TGA Mass Loss and DTG Analyze thermogravimetric mass-loss steps together with the derivative thermogravimetry (DTG) curve. NotyPlot uses the derivative to help resolve loss events and reports step ranges, DTG peak temperatures and residual mass for review.
- Workflow Automatic XRD Peak Analysis in NotyPlot Run a guided automatic XRD peak-analysis pass, then review the detected and fitted results rather than treating automation as a final interpretation. The workflow is designed to shorten setup while keeping peak selection, fitting and scientific checks visible.
- Workflow Calculate Crystallite Size and Microstrain with Williamson-Hall Separate size-related and strain-related contributions to XRD peak broadening with a Williamson–Hall analysis built from several reviewed fitted peaks. The workflow emphasizes peak eligibility and the transformed linear fit because a numerical intercept and slope are only useful when the selected peaks support the model.
- Workflow Calculate Crystallite Size from XRD Using the Scherrer Equation Estimate coherent crystallite size from suitable fitted XRD peak widths using the Scherrer equation. The workflow keeps wavelength, shape factor, fitted FWHM and instrumental broadening visible because each directly affects the reported size.