Documentation

Batch Scale Bars in NotyScale

Prepare multiple images, verify their calibrations and review batch scale bar outputs.

Use the three batch inspector sections

Open Batch and add the intended images. The inspector separates Calibration, Style and Output. Select images deliberately: calibration actions can target the selected set, or missing-calibration images when no selection is active. Review the scope text before applying a saved scale.

1. Review calibrated, pending and missing groups

Calibration groups show the image count and, for verified groups, the unit-per-pixel value. Clicking a group selects its images. Pending detections can be reviewed before acceptance. Use detection or a saved calibration only where appropriate. Manual calibration is available for a single selected image. A pending suggestion should not be treated as verified merely because it is available.

2. Choose a common look and length rule

Style offers the current look where available, a default or larger look, and saved presets. Choose size, anchor and color behavior. Auto length chooses a length per image, while Fixed exposes a numeric distance and unit. Relative presentation can be shared across images while their pixel-to-distance scales remain different. Inspect the result when using automatic color or length choices.

Worked example: three images with different scales

Illustrative inputs: A is 2048 pixels wide at 0.05 micrometers/px, B is 1024 pixels wide at 0.10 micrometers/px, and C is 2048 pixels wide with no verified calibration. A and B each cover 102.4 micrometers across the field, but their pixel scales differ. A fixed 10 micrometer bar spans 200 pixels on A and 100 on B before output transformations. C cannot be given a valid physical bar by copying A simply because it has the same dimensions.

Resolve the uncalibrated image separately

Select C and establish its actual calibration or exclude it from the physical-scale run. Review detections before accepting them. Return to the group counts and verify that the images intended for the run have appropriate calibrations. Reuse the shared visual look on A and B while retaining their independent scales.

The batch Output section offers PNG or TIFF and DPI choices of 150, 300 or 600. Save to library attaches an editable bar to the library image; batch downloads are a ZIP of new output files. Crop footer is a separate option and displays detection counts when enabled. Include report provides an additional record for reviewing the run. Batch controls are distinct from the single-image Export dialog.

Run a representative set and inspect exceptions

Process representative calibrated images first. Compare each output with its expected bar length and check any issues recorded in the report. Inspect footer boundaries and any image that needed an individual edit or reset. Resolve a failed image before rerunning it, rather than repeatedly applying one calibration to the full queue.

Continue your work